Cold vapour atomic fluorescence spectroscopy

Cold vapour atomic fluorescence spectroscopy (CVAFS) is a subset of the analytical technique known as atomic fluorescence spectroscopy (AFS).

Vapour atomic fluorescence analyzer

Use for mercury detection

Used in the measurement of trace amounts of volatile heavy metals such as mercury, cold vapour AFS makes use of the unique characteristic of mercury that allows vapour measurement at room temperature. Free mercury atoms in a carrier gas are excited by a collimated ultraviolet light source at a wavelength of 253.7 nanometres. The excited atoms re-radiate their absorbed energy (fluoresce) at this same wavelength. Unlike the directional excitation source, the fluorescence is omnidirectional and may thus be detected using a photomultiplier tube or UV photodiode.

Gold coated traps may be used to collect mercury in ambient air or other media. The traps are then heated, releasing the mercury from the gold while passing argon through the cartridge. This preconcentrates the mercury, increasing sensitivity, and also transfers the mercury into an inert gas.[1]

Transportable analysers

A number of companies have commercialized mercury detection via CVAFS and produced transportable analysers capable of measuring mercury in ambient air. These devices can measure levels in the low parts per quadrillion range (10−15).

EPA-approved methods

Various analytical methods approved by the United States Environmental Protection Agency (EPA) for measuring mercury in wastewater are in common use. EPA Methods 245.7 and 1631 are commonly used for measurement of industrial wastewater using CVAFS.[2][3]

gollark: Well, this seems like a terrible "solution".
gollark: You didn't say why.
gollark: Who is "they"?
gollark: Poland is not real. Poland does NOT exist.
gollark: As planned.

See also

Other analytical techniques suitable for analyzing heavy metals in air or water:

References

  1. Kopysc, Edyta; Pyrzynska, Krystyna; Garbos, Slawomir; Bulska, Ewa (July 21, 2000). "Determination of Mercury by Cold-Vapor Atomic AbsorptionSpectrometry with Preconcentration on a Gold-Trap". Analytical Sciences. 16 (December 2000): 1309–1312. doi:10.2116/analsci.16.1309. Retrieved April 21, 2016.
  2. "Method 245.7. Mercury in Water by Cold Vapor Atomic Fluorescence Spectrometry; Rev. 2.0". Clean Water Act Analytical Methods. Washington, DC: US Environmental Protection Agency (EPA). February 2005. EPA 821-R-05-001.
  3. "Method 1631. Mercury in Water by Oxidation, Purge and Trap, and Cold Vapor Atomic Fluorescence Spectrometry; Rev. E". Clean Water Act Analytical Methods. EPA. August 2002. EPA 821-R-02-019.
This article is issued from Wikipedia. The text is licensed under Creative Commons - Attribution - Sharealike. Additional terms may apply for the media files.