Flow meter error

In flow measurement, flow meter error is typically reported by a percentage indicating non-linearity of the device. This can be expressed as either a +/- percentage based on either the full range capacity of the device or as a percentage of the actual indicated flow. In practice the flow meter error is a combination of repeatability, accuracy and the uncertainty of the reference calibration. http://www.flowmeters.co.uk/liquid-flow-meter-performance-specification-glossary/

Accuracy of full scale

If a meter’s accuracy is based on a percentage of its full reading capability, the error is a fixed value. For example; an error of 0.5% of full scale, in a 100-gallon per minute (gpm) device is +/- 0.5 gpm. This is the uncertainty all of the time, so as you move away from the full scale capability, the 0.5 gpm error becomes a much larger percentage. At 50 gpm, you are risking a 1% error. At 10 gpm you have a potential 5% error.

Accuracy of reading

If the device has an error expressed as a percentage of the actual flow, then a 0.5% error of 10 gpm is only +/-0.05 gpm; a 10 times better result.[1]