Clifton nanolitre osmometer

A Clifton nanolitre osmometer is a device that allows the determination of the melting and freezing point of an aqueous solution using a sample size of only nanolitres. The device consists of a controller box, a cooling stage, and a sample holder. Additionally, various micrometer syringes, immersion oils and microscopes are required for its use.

Clifton nanolitre osmometer setup

Cooling of the stage is achieved with the use of Peltier devices inside the cooling stage.

The Clifton nanolitre osmometer is especially well suited for determining the antifreeze activity or thermal hysteresis of a solution, which is a difference in the melting and freezing point of a solution. This phenomenon arises when biological antifreeze proteins are present in a solution. Solutions that do not contain antifreeze proteins generally have identical freezing and melting points.

Operation

The cooling stage

A melting and freezing point determination consists of the following steps:

  • Calibration
    • The device should be calibrated with solutions of known osmolarity, typically a 2-point calibration is used.
The sample holder with 7 "wells"
  • Loading
    • Each of the wells in the sample holder are filled with a "heavy" immersion oil.
    • A small amount (generally 1-2nL) of solution is pipetted or blown into the oil in the wells using a micrometer syringe or a pulled capillary tube.
  • Deep Freeze
    • The stage is cooled to a temperature of around -40 degrees Celsius (-40 °F) very quickly (~20 seconds), which causes the samples in the sample wells to flash-freeze.
The Clifton controls and readout
  • Determinations
    • The samples are allowed to melt relatively slowly, under the control of the osmometer controller. The temperature of the sample (solution) at the point when the last ice crystal disappears is called the melting point
    • The samples are refrozen using the deep freeze, and once again melted towards a single ice crystal. When only a single, very small ice crystal remains the samples are cooled in a controlled manner using the nanoliter control box. The temperature of the solution at the point when the ice crystal grows is called the freezing point.

Uses for the Clifton nanolitre osmometer

A scientist using the Clifton
  • Determination of the osmolarity (saltiness) of very small volumes of aqueous solutions
    • Tears
    • Biological samples
    • Plant fluids
  • Determination of the antifreeze activity or thermal hysteresis in a solution.
    • Fluids from polar fishes.
    • Fluids from overwintering insects.
    • Fluids from cold-hardy plants
    • Any fluid that contains antifreeze proteins