Background and Identification

The Nellcor Puritan Bennett NPB-75 is a pulse oximeter and capnograph made by Mallinckrodt’s Nellcor division. Pulse oximeters are non-invasive devices used to monitor a patient’s oxygen saturation. The Nellcor division focuses mainly on producing pulse oximeters and oxygen monitors.

The NPB-75 capnograph/pulse oximeter is used to continuously measure the patient’s end-tidal carbon dioxide, oxygen saturation, pulse rate, and respiratory rate. This monitor is suitable for use with adult, pediatric, and neonatal patients. It is designed for hospital and mobile use, as long as the device is protected from excessive moisture. The NPB-75 is powered either by a rechargeable Nickel Metal Hydride battery back or on AC power.

The NPB-75 capnograph/pulse oximeter is operated using a four-key keypad. The front panel includes an ON/OFF switch, an Alarm Silence/Alarm Menu button, a Contrast/Value Change button, an Event/Home button, and a Next/Menu button.

The model name “NPB-75” is printed in the upper left-hand corner of the monitor’s front face. The monitor’s front panel includes two numeric fields that can display three digits each using a seven-segment LED graphic display. The left-most numeric field displays the end-tidal carbon dioxide. The right-most numeric field displays the patient’s oxygen saturation as a percentage. An LCD graphic display with backlighting presents the patient’s respiratory rate, pulse rate, and a graph of the patient’s oxygen saturation over time.

Technical Specifications

PHYSICAL

  • Size: 206 mm H x 88 mm W x 52.5 mm D (8.11”H x 3.46” W x 2.06”D)
  • Weight: 850 grams (1.87 lb.) (including battery pack)
  • Noise Emission: maximum 45 dB(a)

ENVIRONMENTAL

  • Temperature
    • Operating: 0 oC to 45 oC (32 oF to 113 oF)
    • Storage: -35 oC to 70 oC (-31 oF to 158 oF)
  • Relative Humidity: 10 to 95% (noncondensing)
  • Pressure and Altitude (for operating and storage)
    • Pressure: 430 mmHg to 795 mmHg
    • Altitude: -380m to 4,570m (-1,250 ft. to 15,000 ft.)

SAFETY STANDARDS

  • The monitor was designed to comply with EN60601-1/1990, A1/1993, A2/1995, UL2601-1, CSA C22.2 No. 601.1-M90, ISO 10993-1 (Sensors)

PERFORMANCE

  • Capnograph
    • Sampling Rate: 50 ±7.5 ml/min.
    • CO2 Range: 0-99 mmHg (0-13.2 kPa and 0-13.0 Vol%) at sea level
    • Respiration Rate: 0-150 breaths/min.
    • Warm-up Time: 30 seconds (typical)
    • Rise Time: 190 msec for neonatal mode with low dead space endotracheal tube adapter; 240 msec for adult mode with Microstream Airway Adapter
    • System Response Time: 2.45 seconds (typical), 2.9 seconds maximum (includes delay time and rise time)
    • Ambient Pressure: Compensated internally - automatic
    • Alarms: EtCO2 high, EtCO2 low, FiCO2 high, No Breath.
  • Pulse Oximeter
    • Display Ranges
      • SpO2 Saturation Range: 0-100% SpO2
      • Saturation Accuracy: (% SpO2, ±1 SD)
        • Adults
          • 70-100% ±2 digits
          • 0 - 69% Unspecified
        • Neonates
          • 70-100% ±3 digits
          • 0 - 69% Unspecified
    • Pulse Rate Range: 20-250 beats per minute (bpm)
    • Pulse Rate Accuracy: ± 3 bpm
    • Alarms: SpO2 high, SpO2 low, Pulse high, Pulse low

POWER SPECIFICATIONS

  • External Power Source: 12V DC Medical Grade Adapter
  • Internal Power Source: Ni-MH Rechargeable Battery Pack 7.2V 2100 mA/h (intended for continuous operation)
  • Operating Time (fully charged): Between 3 to 6 hours depending on power management.
  • Recharging Period: Approximately 4.5 hours internal recharging
  • Charger Type: Internal

COMPONENTS AND USER INTERFACE

  • Displays
    • LCD graphic display: (128 x 64 dots) with LED backlight dimension 75 mm x 53 mm.
    • Two numeric fields: 3 digits each, using 7-segment LED graphic display dimension 22mm x 14 mm.
    • Two-color alarm bar: yellow, red
  • Controls and Indicators
    • Front Panel: ON/OFF switch; Alarm Silence/Alarm Menu button; Contrast/Value Change button; Event/Home button; Next/Menu button.
  • Connections
    • Front Panel: CO2 Input connector
    • Top Panel: SpO2 Input socket
    • Rear Panel: Clamp connector, battery release button
    • Side Panel: Power Supply/Communication Adapter Port, Gas output connector